Related Experiment Video
Updated: Aug 12, 2025

Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
Design and performance of a flow sensor CoroQuant used with emergency lung ventilator CoroVent during COVID-19
1Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Czechia.
Abstract:
At the time of COVID-19 pandemic onset in spring 2020 a project CoroVent was initiated with the aim to design and produce emergency lung ventilators and distribute them to hospitals. No flow and tidal volume sensors were available for the project. The lack of tidal volume sensors was a consequence of the rapidly increased demand for mechanical lung ventilators and their consumables. The aim of the study was to develop a special flow sensor CoroQuant for the CoroVent ventilators. The sensor based on pneumotachographic principle, manufactured by the plastic injection moulding of polypropylene, meets the requirements for precision of tidal volume measurement defined by international standard ISO 80601-2-12 for mechanical lung ventilators. CoroVent ventilators with CoroQuant sensors were distributed to 27 hospitals in the Czech Republic for free upon the requests from the medical facilities and started to be clinically used, thus preventing lack of lung ventilators in hospitals in the Czech Republic.
More Related Videos
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
08:26Quantitative Mapping of Specific Ventilation in the Human Lung using Proton Magnetic Resonance Imaging and Oxygen as a Contrast Agent
Published on: June 5, 2019
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation