Measurement of cervical softness before cerclage placement with an aspiration-based device
1Division of Maternal-Fetal Medicine, Department of Obstetrics and Gynecology, Tufts Medical Center, Boston, MA.
American Journal of Obstetrics & Gynecology MFM
|February 1, 2023
Summary
A novel device measured cervical softness, finding that cervices were significantly softer in patients requiring ultrasound-indicated cerclage compared to healthy controls. This technique shows promise for identifying preterm birth risk.
Area of Science:
- Obstetrics and Gynecology
- Maternal-Fetal Medicine
- Biomedical Engineering
Background:
- Cervical insufficiency is linked to preterm birth, with cervical softness playing a role in its pathophysiology.
- Current methods for measuring cervical softness lack routine clinical application.
- Objective measurement of cervical softness could improve the identification of at-risk pregnancies.
Purpose of the Study:
- To measure cervical softness in patients with cervical insufficiency and healthy controls using a novel aspiration-based device.
- To test the hypothesis that the cervix is softer in patients with cervical insufficiency.
Main Methods:
- A cross-sectional study involving 133 women (54 cerclage patients, 79 controls).
- A noninvasive, aspiration-based device measured cervical softness by quantifying aspiration pressure needed for tissue displacement.
- Cerclage patients were categorized into history-indicated, ultrasound-indicated, and examination-indicated groups; controls were matched by gestational age.
Main Results:
- Patients undergoing ultrasound-indicated cerclage exhibited significantly softer cervices (median 62 mbar) compared to second-trimester controls (median 81 mbar; P=.042).
- No significant differences in cervical softness were found between history-indicated or examination-indicated cerclage groups and their respective controls.
- Data from 133 women were analyzed, with 54 in the cerclage group and 79 controls.
Conclusions:
- An aspiration-based device demonstrated that cervices are significantly softer in patients presenting for ultrasound-indicated cerclage than in normal controls.
- Quantitative measurement of cervical softness using this device is a promising method for objective assessment during pregnancy.
- This technique may aid in identifying pregnancies at risk for cervical insufficiency and subsequent preterm birth.