SYRINGE DESIGN AND FILLING TECHNIQUE AFFECT ACCURACY OF ANTI-VASCULAR ENDOTHELIAL GROWTH FACTOR INTRAVITREAL
Mark Krauthammer1, Gal Harel2, Elad Moisseiev2,3
1Department of Ophthalmology, Tel Aviv Medical Center, Tel Aviv, Israel.
Purpose:
To evaluate the effect of syringe design and filling technique on the accuracy of anti-vascular endothelial growth factor delivery.
Methods:
Volume output was measured with three syringe designs: a 1.0-mL slip-tip syringe, a 1.0-mL Luer-lock syringe, and a ranibizumab prefilled syringe-using two filling techniques ("upward" and "downward") and two fluids (water and bevacizumab). A total of 300 simulated injections were performed. Accuracy was determined by difference from the intended volume of 50 µ L and by mean absolute percentage error.
Results:
Volume outputs were significantly different between syringe designs, with mean values of 61.99 ± 4.18 µ L with the 1-mL slip-tip syringe, 57.43 ± 4.95 µ L with the Luer-lock 1-mL syringe, and 51.06 ± 4.74 µ L with the ranibizumab syringe, making the latter the most accurate syringe. There were 37 cases (12.3%) of underdosing below 50 µ L, the majority of which occurred with the ranibizumab syringe. The "downward" technique reduced the occurrence of air bubbles.
Conclusion:
Intravitreal injections using 1.0-mL syringes are less accurate than using the ranibizumab prefilled syringe, which has a low-volume and low dead-space plunger design. The variability in volume output may result in less predictable treatment response, especially in cases of underdosing, which were more common with the ranibizumab syringe.