Angiojet™ mechanical thrombectomy-induced hemolysis triggering blood-leak alarms
Babitha Selvananthan1, Omar Aziz2, Arvind D Lee2
1Nepean Kidney Research Centre, Department of Renal Medicine, Nepean Hospital, Kingswood, Australia.
Abstract:
Blood leak alarms are increasingly rare and may be triggered by rupture of the dialyzer membrane, or intravascular hemolysis. We report two patients who developed hemolysis following Angiojet™ thrombolysis and thrombectomy of thrombosed arteriovenous fistulas, triggering blood leak alarms on hemodialysis, the occurrence of which has not been reported before. AngioJet™-induced hemolysis should be considered in the differential diagnosis for blood leak alarms occurring soon after an intervention on an arterio-venous (AV) fistula.