Ventricular arrhythmia detection for contemporary Biotronik and Abbott implantable cardioverter defibrillators with
Adam Oesterle1, Sanket S Dhruva2, Cara N Pellegrini2
1Division of Cardiology, Department of Medicine, University of California San Francisco - Veterans Affairs San Francisco Health Care System, 4150 Clement Street, Building 203, Room 2A-25, San Francisco, CA, 94121, USA. adam.oesterle@ucsf.edu.
Background:
Implantable cardioverter defibrillators (ICDs) are typically programed with both ventricular tachycardia (VT) and ventricular fibrillation (VF) treatment zones. Biotronik and Abbott ICDs do not increment the VT counter when the tachycardia accelerates to the VF zone, which could result in a prolonged delay in tachycardia detection.
Methods:
Patients with Biotronik and Abbott ICDs receiving care at Veterans Affairs facilities in Northern California were identified. Patient information and device tracings for patients with any ICD therapies were examined to assess for possible delayed tachycardia detection.
Results:
Among 52 patients with Biotronik ICDs, 8 (15%) experienced appropriate ICD therapy over a median follow-up of 29 months. Among 68 patients with Abbott ICDs, 26 (38%) experienced appropriate ICD therapy over a median follow-up of 83 months. Three of the patients with Biotronik ICDs who received appropriate therapy experienced a delay in VT/VF detection due to the tachycardia rate oscillating between the VT and VF treatment zones (longest 31.2 s on detection), compared with four of the patients with Abbott ICDs (longest 4.1 s on the detection and 8 s on redetect). One of the patients with a Biotronik ICD experienced recurrent syncope associated with delayed detection and another died on the day of delayed detection. One of the patients with an Abbott ICD experienced syncope.
Conclusions:
Because contemporary Biotronik and Abbott ICDs freeze the VT counters when tachycardia is in the VF zone, ICD therapies can be markedly delayed when the tachycardia oscillates between the VT and VF zone.
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias VI: Management of Dysrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Mechanism of Cardiac Arrhythmias