Investigation of in-line filter replacement intervals for infusion
H Kato1, Y Morikawa2, M Hagihara3
1Department of Pharmacy, Mie University Hospital, Mie, Japan; Department of Clinical Pharmaceutics, Division of Clinical Medical Science, Mie University Graduate School of Medicine, Mie, Japan; Department of Clinical Infectious Diseases, Aichi Medical University, Aichi, Japan.
Background:
In-line filters in peripheral and central venous catheters are used to remove bacterial cells mechanically. A recent study indicated an extension of the use of infusion sets to 7 days. There is no evidence regarding replacement intervals for in-line filters.
Aim:
To test in-line filters that were used continuously for 7 days in order to investigate their ability to remove bacteria and assess the flow rate.
Methods:
Three different in-line filters were attached to an ELNEOPA-NF No. 2 premixed infusion bag of intravenous hyperalimentation, into which Staphylococcus epidermidis ATCC12228 or Escherichia coli ATCC25922 was inoculated. These experiments were compared with a control infusion. The infusion was dropped at a flow rate of 40 mL/h and replaced at 24-h intervals for 7 days. Samples were collected 24 h after drop initiation.
Findings:
S. epidermidis was not detected in droplets between Days 1 and 6, but In-line filters 1 and 2 showed droplets containing 6-10 colony-forming units/mL on Day 7. E. coli was not detected in any of the filters after 7 days of continuous use. Flow rates <40 mL/h were observed on Day 7 for In-line filter 3 in studies of S. epidermidis, and on Days 4 and 3 for In-line filters 2 and 3, respectively, in studies of E. coli.
Conclusion:
This study revealed differences in bacterial removal and flow rates under high inoculation between the three in-line filters tested. It is suggested that in-line filters can be used continuously for a maximum of 6 days, and reductions in flow rate after 48 h of continuous use should be noted carefully.
More Related Videos
12:11A Modified EPA Method 1623 that Uses Tangential Flow Hollow-fiber Ultrafiltration and Heat Dissociation Steps to Detect Waterborne Cryptosporidium and Giardia spp.
Published on: July 9, 2012
08:48Analysis of the Ambient Particulate Matter-induced Chromosomal Aberrations Using an In Vitro System
Published on: December 21, 2016
Related Concept Videos
Filtration
Continuous Renal Replacement Therapy
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Dialysis
Endocarditis IV: Nursing Management