An alarm device for mechanical compression device displacement at femoral artery puncture sites
Yue-Huan Gu1, Shu-Jin Ye1, Li-Wei Song1
1Department of Nursing, Renji Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai, China.
Objective:
To develop an alarm device for the mechanical compression device displacement (MCD), and further evaluate its effectiveness in clinical use.
Material And Methods:
The alarm device is mainly composed of buzzer, indicator light, magnetic sheet. This is a prospective randomized and controlled study. Four hundred patients who met the inclusion/exclusion criteria were included and randomly assigned to two groups (MCD group vs alarm + MCD group). The primary outcome measures were the sensitivity and specificity of the alarm device to detect MCD displacement, time to hemostasis (TTH), time to ambulation (TTA), time to hospital discharge (TTHD), hospital costs (HC), complication rates, and patient satisfaction.
Results:
The sensitivity and specificity of the alarm device in detecting MCD displacement were 94.44% and 88.46%, respectively. The study group achieved shorter TTH (p = .034), shorter TTA (p = .021), lower complication rates (p = .025), and better patients' satisfaction (p < .001) compared to the control group. However, no significant difference was observed in TTHD (p = .361) and HC (p = .583).
Conclusion:
The alarm device is highly sensitive in detecting MCD displacement, while achieving better clinical outcomes compared with artificial monitoring.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure in the Leg
Preparation:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse Assessment Sites
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure