Correction of the urinary testosterone to epitestosterone ratio measurement in antidoping analyses by chromatographic and mass spectrometric techniques
Related Experiment Video
Updated: Aug 10, 2025

Deep Proteome Profiling by Isobaric Labeling, Extensive Liquid Chromatography, Mass Spectrometry, and Software-assisted Quantification
Published on: November 15, 2017
Dayamín Martínez Brito1,2, Teresa Correa Vidal1, Xavier de la Torre2
1Laboratorio Antidoping de la Habana, Analytical Department, Havana, Cuba, Cuba.
Abstract:
The ratio of testosterone (T) to epitestosterone (E) determined in urine samples is the main biomarker used to prove T and precursors abused by athletes. Analytically, the correction of the ratio is required by the World Anti-Doping Agency. This work describes a series of experiments aimed to study when it is appropriate to correct the T/E ratio value, using different mass spectrometric techniques since only reports on GC-MS exist. Analyses using external calibrators, controls, and routine samples were performed by three different techniques GC-MS, GC-MSn , and LC-MSn . A statistical comparison of the T/E was performed after the application of two corrections previously published: Isotopic contribution peak area correction (corr_1) and use of a verified internal deuterated internal standard (TD3/ED3) correction (corr_2) and the ratio based on T and E concentrations. The use of external calibration samples introduces biases that influence not only the T and E concentrations but also the T/E ratio, even when both methods of correction are applied. The correction after applying corr_1 method barely contributed to the accuracy of the T/E calculation. Nevertheless, the application of the corr_2 method increased the accuracy between 5% and 6% when comparing theoretical and experimental T/E values. Finally, the best results were obtained by the ratio calculated directly from the estimated concentrations of T and E. Attention must be paid when the T/E is measured by LC-MSn since different acquisition modes produced significantly different results, even in MS/MS when the same transition is used for T and E.
More Related Videos
08:41Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019