A Novel Locking Buttress Plate Designed for Simultaneous Medial and Posterolateral Tibial Plateau Fractures: Concept
Bangji Yan1, Xiaotao Huang1, Yingxing Xu2
1Department of Orthopaedics and Traumatology, Cixi Hospital of Traditional Chinese Medicine, Ningbo, China.
Objective:
The complex tibial plateau fractures involving both medial and posterolateral columns are of frequent occurrence in clinics, but the existing fixation system cannot deal with medial and posterolateral fragments simultaneously. Therefore, a novel locking buttress plate named as medial and posterior column plate (MPCP) was designed in this study to fix the simultaneous medial and posterolateral tibial plateau fractures. Meanwhile, the comparative finite element analysis (FEA) was conducted to investigate the discrepancy between MPCP and traditional multiple plates (MP + PLP) in their biomechanical characteristics.
Methods:
Two 3D finite element models of simultaneous medial and posterolateral tibial plateau fracture fixed with MPCP and MP + PLP system, respectively, was constructed. To imitate the axial stress of knee joint in ordinary life, diverse axial forces with 100, 500, 1000, and 1500 N were applied in the two fixation models, and then the equivalent displacement and stress nephograms and values were obtained.
Results:
The similar trend of displacement and stress increasing with the loads was observed in the two fixation models. However, several heterogeneities of displacement and stress distribution were found in the two fixation models. The max displacement and von Mises stress values of plates, screws, and fragments in the MPCP fixation model were significantly smaller than that in the MP + PLP fixation model, except for the max-shear stress values.
Conclusion:
As a single locking buttress plate, the MPCP system showed the excellent benefit on improving the stability of the simultaneous medial and posterolateral tibial plateau fractures, compared with the traditional double plate fixation system. However, the excessive shear stress around screw holes should be paid attention to prevent trabecular microfracture and screw loosening.
More Related Videos
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
07:35Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
Related Concept Videos
Bones of the Lower Limb: Tibia and Fibula
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...