Five-year evaluation of a low-cost quality assurance protocol for clinical ultrasound transducers
Niels Hyldgaard1, Lise Bolander Malvang1, Lau Brix2
1Department of Radiology, Diagnostic Centre, University Research Clinic for Innovative Patient Pathways, Silkeborg Regional Hospital, Silkeborg, Denmark.
Objective:
To implement, execute and evaluate a quality assurance test protocol for clinical ultrasound transducers that is software-based, user-independent, inexpensive, easily repeatable, and does not require tissue phantoms.
Methods:
The test protocol is based on in-air reverberation images. The software test tool generates uniformity- and reverberation profiles to monitor system sensitivities and signal uniformities that provide a sensitive analysis of transducer status. Whenever a transducer was suspected to be damaged validating tests were conducted using the Sonora FirstCall test system. A total of 21 transducers from five ultrasound scanner systems were included in the study. Tests were performed every other month for a total of 5 years.
Results:
Each transducer was tested on average 11.7 times. Testing a transducer required a total of 2.75 hours annually. The ultrasound quality assurance test protocol detected an average annual fail rate of 10.7%. The test protocol provides a reliable approach for monitoring transducer lens status of clinically used ultrasound transducers.
Conclusion:
The ultrasound quality assurance test protocol may potentially find deviations in diagnostic quality before being noticed by the clinicians. Thus, the ultrasound quality assurance test protocol holds the ability to reduce the risk of undiscovered image quality degradation hereby minimizing the danger of diagnostic errors.
More Related Videos
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
16:01An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called...