Comparison of Autof Ms1000 and EXS3000 MALDI-TOF MS Platforms for Routine Identification of Microorganisms
Lijuan Xiong1, Xu Long1, Lijun Ni2
1Department of Laboratory Medicine, the Second Affiliated Hospital of Guizhou University of Traditional Chinese Medicine, Guizhou, People's Republic of China.
Purpose:
Matrix-assisted laser desorption-ionization-time of flight mass spectrometry (MALDI-TOF) has recently been widely used in clinical microbiology laboratories, with the advantages of being reliable, rapid, and cost-effective. Here, we reported the performance of two MALDI-TOF MS instruments, EXS3000 (Zybio, China) and Autof ms1000 (Autobio, China), which are commonly used in clinical microbiology field.
Methods:
A total of 209 common clinical common isolates, including 70 gram-negative bacteria strains, 58 gram-positive bacteria strains, 33 yeast strains, 15 anaerobic bacteria strains, and 33 mold strains, and 19 mycobacterial strains were tested. All strains were identified by EXS3000 (Zybio, China) and Autof ms1000 (Autobio, China). Sequence analysis of 16S rRNA or ITS regions was used to verify all strains.
Results:
Current study found that species-level discrimination was found to be 191 (91.39%) and 190 (90.91%) by EXS3000 and Autof ms1000, respectively. Genus-level discrimination was 205 (98.09%) by the EXS3000 and 205 (98.09%) by the Autof ms1000, respectively. The correct results at species level of the EXS3000 were 91.43% (64/70) for gram-negative bacteria, 93.1% (54/58) for gram-positive cocci, 93.94% (31/33) for yeast, 100% (15/15) for anaerobes and 81.82% (27/33) for filamentous fungi. The correct results at species level of the Autof ms1000 were 92.86% (65/70) for gram-negative bacteria, 91.38% (53/58) for gram-positive cocci, 93.94% (31/33) for yeast, 100% (15/15) for anaerobes and 78.79% (26/33) for filamentous fungi.
Conclusion:
Although the results show that the EXS3000 and Autof ms1000 systems are equally good choices in terms of analytical efficiency for routine procedures, the test result of EXS3000 is slightly better than Autof ms1000. It's worth mentioning that the target plate of the EXS 3000 instrument is reusable, but the target plate of the Autof ms1000 is disposable, making the EXS3000 more effective in reducing costs.
More Related Videos
13:29Rapid Identification of Gram Negative Bacteria from Blood Culture Broth Using MALDI-TOF Mass Spectrometry
Published on: May 28, 2014
09:26Identification of Antibacterial Immunity Proteins in Escherichia coli using MALDI-TOF-TOF-MS/MS and Top-Down Proteomic Analysis
Published on: May 23, 2021
Related Concept Videos
MALDI-TOF Mass Spectrometry
Matrix-assisted laser desorption ionization (MALDI) is a commonly...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Matrix-Assisted Laser Desorption Ionization (MALDI)
The analyte of interest, a biomolecule or a mixture of biomolecules, is mixed with a suitable matrix material. The...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...