Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Severe Hypercapnia during Anaesthesia under Mechanical Ventilation in Two Paediatric Patients
Anastasia Papastefanou1, Eva Rioja1
1Optivet Referrals, Ltd., Havant PO9 2NJ, UK.
Abstract:
A 2-month-old male 1.56 kg Yorkshire terrier (Case No. 1) and a 3-month-old male 2.3 kg Jack Russell Terrier (Case No. 2) were scheduled for ophthalmological surgery under general anaesthesia and neuromuscular blockade. For both patients, volume-controlled ventilation (VCV) was used with set tidal volumes (VT) of 13 mL/kg and 20 mL/kg for cases No. 1 and 2, respectively. The type of ventilator used did not take into account the intrinsic compliance of the breathing system; therefore, a significant part of the delivered VT was wasted in the expansion of the breathing system, and did not reach the patients, causing alveolar hypoventilation. Both cases developed low dynamic compliance (CD), and after a recruitment manoeuvre, EtCO2 of up to 116 mmHg and 197 mmHg were revealed for cases No. 1 and 2, respectively. The two cases had to be ventilated manually, using positive inspiratory pressures (PIP) of 20-25 mmHg, in order to improve alveolar ventilation and reduce the EtCO2, as adjustments to the VCV were ineffective. Both patients maintained an oxygen haemoglobin saturation between 94% and 100% throughout the procedure and they recovered well. Using a higher VT from the beginning, to compensate for the compliance of the breathing system, or the use of pressure-controlled ventilation (PCV), could have potentially helped to avoid these two incidences of severe hypercapnia.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Acute Respiratory Failure-III
Hyperpnea and Hyperventilation
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...
Mechanical Ventilation I: Indication and Settings
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include: