Non-invasive assessment of Pulse Wave Transit Time (PWTT) is a poor predictor for intraoperative fluid
Kimiko Fukui1, Johannes M Wirkus1, Erik K Hartmann1
1Department of Anesthesiology, University Medical Center of the Johannes Gutenberg-University, Langenbeckstraße 1, 55131, Mainz, Germany.
Background:
Aim of this study is to test the predictive value of Pulse Wave Transit Time (PWTT) for fluid responsiveness in comparison to the established fluid responsiveness parameters pulse pressure (ΔPP) and corrected flow time (FTc) during major abdominal surgery.
Methods:
Forty patients undergoing major abdominal surgery were enrolled with continuous monitoring of PWTT (LifeScope® Modell J BSM-9101 Nihon Kohden Europe GmbH, Rosbach, Germany) and stroke volume (Esophageal Doppler Monitoring CardioQ-ODM®, Deltex Medical Ltd, Chichester, UK). In case of hypovolemia (difference in pulse pressure [∆PP] ≥ 9%, corrected flow time [FTc] ≤ 350 ms) a fluid bolus of 7 ml/kg ideal body weight was administered. Receiver operating characteristics (ROC) curves and corresponding areas under the curve (AUCs) were used to compare different methods of determining PWTT. A Wilcoxon test was used to discriminate fluid responders (increase in stroke volume of ≥ 10%) from non-responders. The predictive value of PWTT for fluid responsiveness was compared by testing for differences between ROC curves of PWTT, ΔPP and FTc using the methods by DeLong.
Results:
AUCs (area under the ROC-curve) to predict fluid responsiveness for PWTT-parameters were 0.61 (raw c finger Q), 0.61 (raw c finger R), 0.57 (raw c ear Q), 0.53 (raw c ear R), 0.54 (raw non-c finger Q), 0.52 (raw non-c finger R), 0.50 (raw non-c ear Q), 0.55 (raw non-c ear R), 0.63 (∆ c finger Q), 0.61 (∆ c finger R), 0.64 (∆ c ear Q), 0.66 (∆ c ear R), 0.59 (∆ non-c finger Q), 0.57 (∆ non-c finger R), 0.57 (∆ non-c ear Q), 0.61 (∆ non-c ear R) [raw measurements vs. ∆ = respiratory variation; c = corrected measurements according to Bazett's formula vs. non-c = uncorrected measurements; Q vs. R = start of PWTT-measurements with Q- or R-wave in ECG; finger vs. ear = pulse oximetry probe location]. Hence, the highest AUC to predict fluid responsiveness by PWTT was achieved by calculating its respiratory variation (∆PWTT), with a pulse oximeter attached to the earlobe, using the R-wave in ECG, and correction by Bazett's formula (AUC best-PWTT 0.66, 95% CI 0.54-0.79). ∆PWTT was sufficient to discriminate fluid responders from non-responders (p = 0.029). No difference in predicting fluid responsiveness was found between best-PWTT and ∆PP (AUC 0.65, 95% CI 0.51-0.79; p = 0.88), or best-PWTT and FTc (AUC 0.62, 95% CI 0.49-0.75; p = 0.68).
Conclusion:
ΔPWTT shows poor ability to predict fluid responsiveness intraoperatively. Moreover, established alternatives ΔPP and FTc did not perform better.
Trial Registration:
Prior to enrolement on clinicaltrials.gov (NC T03280953; date of registration 13/09/2017).
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring pulse
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: