Influence of saturation effects on biexponential liver intravoxel incoherent motion
Martin Loh1, Tobit Führes1, Christoph Stuprich1
1Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
Purpose:
Studies on intravoxel incoherent motion (IVIM) imaging in the liver have been carried out with different acquisition protocols. The number of acquired slices and the distances between slices can influence IVIM measurements due to saturation effects, but these effects have often been disregarded. This study investigated differences in biexponential IVIM parameters between two slice settings.
Methods:
Fifteen healthy volunteers (21-30 years) were examined at a field strength of 3 T. Diffusion-weighted images of the abdomen were acquired with 16 b values (0-800 s/mm2 ), with four slices for the few slices setting and 24-27 slices for the many slices setting. Regions of interest were manually drawn in the liver. The data were fitted with a monoexponential signal curve and a biexponential IVIM curve, and biexponential IVIM parameters were determined. The dependence on the slice setting was assessed with Student's t test for paired samples (normally distributed IVIM parameters) and the Wilcoxon signed-rank test (non-normally distributed parameters).
Results:
None of the parameters were significantly different between the settings. For few slices and many slices, respectively, the mean values (SDs) for were ( ) and ( ); for they were 29.7% (6.2%) and 27.7% (3.6%); and for they were ( ) and ( ).
Conclusion:
Biexponential IVIM parameters in the liver are comparable among IVIM studies that use different slice settings, with mostly negligible saturation effects. However, this may not hold for studies that use much shorter TR.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Drug Concentration Versus Time Correlation
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the...
Imaging Studies VII: Vascular Imaging
Noncompartmental Analysis: Mean Transit, Absorption and Dissolution Time
One of the key parameters is the mean transit time (MTT), which refers to the total duration required for drug molecules to transit through the body. MTT is determined by calculating the ratio of the area under the moment curve to the area...