Related Experiment Video
Updated: Aug 8, 2025

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Effect of Implant Linkage on Axial and Rotational Stiffness of Nail-Plate Constructs for Comminuted Distal Femoral
Charles C Lin1, Nicholas Parody, Utkarsh Anil
1NYU Langone Orthopedic Hospital, NYU Langone Health, New York, NY.
Objectives:
To determine the effect of linking the intramedullary nail and the laterally applied locking plate to treat comminuted distal femur fractures and allow for immediate weight bearing.
Methods:
Comminuted extra-articular distal femur fractures were created in 16 synthetic osteoporotic femurs and split into 2 groups: linked versus unlinked. In the linked construct, in addition to standard plate fixation and proximal locking of the nail, 2 nonthreaded locking bolts (prototypes) were placed through both the plate and the nail. In the unlinked construct, the same number of screws affixed the plate to the bone but were directed around the nail, and separate distal interlocking screws were placed for nail fixation. Each specimen underwent sequential axial and torsional loading, and axial and torsional stiffness were calculated and compared.
Results:
On average, the unlinked constructs demonstrated a greater axial stiffness at all levels of axial loading, and linked constructs demonstrated greater average rotational stiffness. However, there were no statistically significant differences ( P > 0.189) between the linked and unlinked groups at any axial or torsional load.
Conclusion:
In distal femur fractures with metaphyseal comminution, there were no significant differences in axial or torsional stiffness when linking the plate to the nail. Linking the construct seems to provide no significant mechanical advantage compared with the unlinked environment, but it may be a useful strategy to reduce nail "traffic" in the distal segment with no significant disadvantage.
More Related Videos
04:42Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018