Related Experiment Video
Updated: Jul 3, 2026

Tissue-simulating Phantoms for Assessing Potential Near-infrared Fluorescence Imaging Applications in Breast Cancer Surgery
Published on: September 19, 2014
Identifying Parathyroids in Pediatric Thyroid/Parathyroid Surgery by Near Infrared Autofluorescence
Ryan H Belcher1,2, Giju Thomas3,4, Parker A Willmon3,4
1Vanderbilt Department of Otolaryngology - Head and Neck Surgery, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
Objectives:
Compared to adult patients undergoing thyroid surgery, pediatric patients have higher rates of hypoparathyroidism often related to parathyroid gland (PG) inadvertent injury or devascularization. Previous studies have shown that near-infrared-autofluorescence (NIRAF) can be reliably used intraoperatively for label-free parathyroid identification, but all prior studies have been performed in adult patients. In this study, we assess the utility and accuracy of NIRAF with a fiber-optic probe-based system to identify PGs in pediatric patients undergoing thyroidectomy or parathyroidectomy.
Methods:
All pediatric patients (under 18 years of age) undergoing thyroidectomy or parathyroidectomy were enrolled in this IRB-approved study. The surgeon's visual assessment of tissues was first noted and the surgeon's confidence level in the tissue identified was recorded. A fiber-optic probe was then used to illuminate tissues-of-interest with a wavelength of 785 nm and resulting NIRAF intensities from these tissues were measured while the surgeon was blinded to results.
Results:
NIRAF intensities were measured intraoperatively in 19 pediatric patients. Normalized NIRAF intensities for PGs (3.63 ± 2.47) were significantly higher than that of thyroid (0.99 ± 0.36, p < 0.001) and other surrounding soft tissues (0.86 ± 0.40, p < 0.001). Based on the PG identification ratio threshold of 1.2, NIRAF yielded a detection rate of 95.8% (46/48 pediatric PGs).
Conclusion:
Our findings indicate that NIRAF detection can potentially be a valuable and non-invasive technique to identify PGs during neck operations in the pediatric population. To our knowledge, this is the first study in children to assess the accuracy of probe-based NIRAF detection for intraoperative parathyroid identification.
Level Of Evidence:
Level 4 Laryngoscope, 133:3208-3215, 2023.