Related Experiment Video
Updated: Aug 7, 2025

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Sigmoid dose accumulation and reporting for multifractionated brachytherapy for cervical cancer: Methodological
Arpita Bindal1, Prachi Mittal1, Abhishek Shinghal1
1Department of Radiation Oncology and Medical Physics, Tata Memorial Hospital, Tata Memorial Centre, Homi Bhabha National Institute, Mumbai, Maharashtra, India.
Purpose:
The sigmoid is an important organ at risk for gynecological brachytherapy (BT). However, the reliability of localization of high-dose regions during multi-fractionated treatment is limited. This work reports the methodological development of sigmoid points to summate multi-fractionated doses.
Methods And Material:
Ten paired MRI data sets of ring-based intracavitary brachytherapy were obtained. Simulating a virtual endoscope, a reference line was created along the central axis of the anorectosigmoid for each implant. A trendline was generated, and linear dose was determined. Three-dimensional (3D) coordinates of high-dose regions were identified, and overlap was determined. In the next step, 3D coordinates of high-dose sigmoid points were localized in reference to cervical os and re-verified for location in reference to sigmoid lumen and corroboration with 2cc doses. With minor modifications, sigmoid points were proposed.
Results:
In 6 of 10 patients, high-dose regions co-localized in subsequent fractions of BT. Three high-dose regions were identified along the sigmoid length and proposed as sigmoid points in reference to cervical os. (S1'= 0.5 cm right, 1.5 cm posterior, and 2.4 cm cranial; S2' = 0.3 cm anterior and 4.5 cm cranial; S3' = 2.7 cm left, 3 cm anterior, and 3.6 cm cranial to the cervical os). S1' and S2' were located in the sigmoid in 70% and 60% of data sets. The mean difference between D2cc and S1'/S2' was 0.30 Gy and 1.06 Gy respectively. S3' had limited corroboration to sigmoid lumen or 2 cc doses. The points S1' and S2' were further modified (minor) for applicability and proposed as sigmoid points 1 and 2 (SP1 0.5 right,1.5 posterior and 2.5 cm cranial to cervical os and SP2 (0.5 cm anterior and 4.5 cm cranial to cervical os)).
Conclusion:
SP1 and SP 2 are proposed as a surrogate for 2 cc sigmoid doses and may provide a method of reliable inter-fraction dose summation. This pilot work requires further validation.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019