A Framework for ExacTrac Dynamic Commissioning for Stereotactic Radiosurgery and Stereotactic Ablative Radiotherapy
Ben Perrett1, Jaysree Ukath1, Emma Horgan1
1Department of Radiation Oncology, Princess Alexandra Hospital, Woolloongabba, Queensland, Australia.
Journal of Medical Physics
|March 13, 2023
Summary
This study provides a framework for commissioning ExacTrac Dynamic image-guided and surface-guided radiotherapy (SGRT) systems. All tests met established tolerances, ensuring accurate patient treatment with SGRT and X-ray image fusion.
Area of Science:
- Medical Physics
- Radiotherapy Technology
- Image Guidance
Background:
- Image-guided radiotherapy (IGRT) and surface-guided radiotherapy (SGRT) are crucial for precise radiation delivery.
- The ExacTrac Dynamic system integrates advanced imaging and surface tracking for enhanced accuracy.
- Commissioning is essential to validate system performance and ensure patient safety.
Purpose of the Study:
- To establish a comprehensive framework for commissioning tests and tolerances for the ExacTrac Dynamic IGRT and SGRT system.
- To detail the commissioning process and results for a newly installed ExacTrac Dynamic system on an Elekta Versa HD linear accelerator.
- To provide guidance for validating system performance in various clinical workflows.
Main Methods:
- Developed and implemented a suite of commissioning tests covering safety, calibration, image quality, SGRT performance, and data integrity.
- Utilized custom phantom designs for evaluating deep inspiration breath-hold (DIBH), implanted markers, and gating workflows.
- Assessed X-ray image fusion reproducibility and SGRT localization accuracy.
Main Results:
- All commissioning tests were successfully completed within established tolerances.
- Maximum translational and rotational deviations for SGRT were 0.3 mm and 0.4°, respectively.
- Standard deviation for X-ray image fusion reproducibility was 0.08 mm.
Conclusions:
- The ExacTrac Dynamic system meets performance specifications for image-guided and surface-guided radiotherapy.
- The established commissioning framework and demonstrated results support the clinical implementation of this system.
- Ongoing validation and workflow-specific testing are recommended for continuous quality assurance.