Assessment of the Androstenedione Assay on a Roche Analytical Platform and Comparison with Different Methods
Clinical Laboratory
|March 13, 2023
Summary
The new Roche® assay for D4-androstenedione (D4ASD) shows acceptable analytical performance for diagnosing androgenic hyperproduction. Its precision and accuracy align with established standards, offering a reliable alternative to existing methods.
Area of Science:
- Endocrinology
- Clinical Chemistry
- Analytical Chemistry
Background:
- D4-androstenedione (D4ASD) is a key intermediate in androgen biosynthesis.
- Accurate D4ASD measurement is crucial for diagnosing androgenic hyperproduction disorders.
- Existing quantification methods, particularly immunoassays, face challenges with steroid cross-reactivity.
Purpose of the Study:
- To evaluate the analytical performance of a new, fully automated D4ASD assay from Roche®.
- To compare the Roche® D4ASD assay with radioimmunoassay (RIA) and liquid chromatography-tandem mass spectrometry (LC-MS/MS).
Main Methods:
- Assessment of repeatability and intermediate precision for the D4ASD Roche® assay.
- Comparison of the Roche® assay against RIA and LC-MS/MS methods.
- Evaluation based on the RICOS standard and correlation analysis (slope, y-intercept).
Main Results:
- The D4ASD Roche® assay demonstrated acceptable repeatability and intermediate precision (CV ≤ 7.9%) per the RICOS standard.
- The assay showed good correlation with reference methods, with acceptable 95% confidence intervals for slope and y-intercept.
- Analytical performance criteria, including imprecision and trueness, met predefined acceptance limits.
Conclusions:
- The Roche® automated assay for D4-androstenedione (D4ASD) exhibits acceptable analytical performance.
- The method provides reliable results with intermediate imprecision and trueness within acceptable limits.
- This assay presents a viable option for D4ASD quantification in clinical diagnostics.