Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
[Dosimetric Properties of Brass Mesh Bolus for High-energy Photon Beam]
Hikari Hara1, Yusuke Shibayama1, Jun-Ichi Fukunaga1
1Division of Radiology, Department of Medical Technology, Kyushu University Hospital.
Purpose:
To investigate fundamental dosimetric properties of surface dose, exit dose, and beam profile of the brass mesh bolus for 4, 6, and 10 MV high-energy photon beams in radiation therapy.
Methods:
Surface dose and exit dose in the water-equivalent phantom were measured, and percent depth doses (PDDs) were calculated with no bolus, one layer of brass mesh, two layers of brass mesh bolus, three layers of brass mesh bolus, and 0.5 cm tissue-equivalent (TE) bolus. Exit dose was measured at a phantom thickness of 10 cm. Beam profiles were measured at phantom depths of 0 cm and 10 cm. All dosimetry was performed for 4, 6, and 10 MV photon beams using a linear accelerator.
Results:
The surface dose at a phantom depth of 0 cm increased to 37.3%, 36.3%, and 31.0% for 4, 6, and 10 MV, respectively, with the brass mesh bolus compared to the case of no bolus. The surface dose decreased with one layer of brass mesh bolus compared to that with the 0.5 cm TE bolus. On the other hand, the exit dose increased to 22.0%, 23.1%, and 22.8% for 4, 6, and 10 MV, respectively, with the brass mesh bolus compared to the case of no bolus. The beam profile at the depth of 0 cm showed oscillations, and the difference between the maximum and minimum doses was up to 13.1% with one layer of brass mesh bolus.
Conclusion:
It was suggested that the brass mesh bolus not only increases the surface dose but also has different properties from the conventional TE bolus.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force...
Radiation: Applications
The average...
Dual Nature of Electromagnetic (EM) Radiation
Wavelength is the distance between two consecutive peaks (the highest point) or troughs (the lowest point) in the wave. Frequency is the...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Absorption of Radiation