Monte Carlo based continuous aperture optimization for VMAT on conventional and MR-linacs: A proof of concept
Shiqin Su1, Parmveer Atwal2, I Antoniu Popescu3,4
1Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan, USA.
Background:
Currently, the commercial treatment planning systems for magnetic-resonance guided linear accelerators (MR-linacs) only support step-and-shoot intensity-modulated radiation therapy (IMRT). However, recent studies have shown the feasibility of delivering arc therapy on MR-linacs, which is expected to improve dose distributions and delivery speed. By accurately accounting for the electron return effect in the presence of a magnetic field, a Monte Carlo (MC) algorithm is ideally suited for the inverse treatment planning of this technique.
Purpose:
We propose a novel MC-based continuous aperture optimization (MCCAO) algorithm for volumetric modulated arc therapy (VMAT), including applications to VMAT on MR-linacs and trajectory-based VMAT. A unique feature of MCCAO is that the continuous character of gantry rotation and multileaf collimator (MLC) motion is accounted for at every stage of the optimization.
Methods:
The optimization process uses a multistage simulation of 4D dose distribution. A phase space is scored at the top surface of the MLC and the energy deposition of each particle history is mapped to its position in this phase space. A progressive sampling method is used, where both MLC leaf positions and monitor unit (MU) weights are randomly changed, while respecting the linac mechanical limits. Due to the continuous nature of the leaf motion, such changes affect not only a single control point, but propagate to the adjacent ones as well, and the corresponding dose distribution changes are accounted for. A dose-volume cost function is used, which includes the MC statistical uncertainty.
Results:
We applied our optimization technique to various treatment sites, using standard and flattening-filter-free (FFF) 6 MV beam models, with and without a 1.5 T magnetic field. MCCAO generates deliverable plans, whose dose distributions are in good agreement with measurements on ArcCHECK and stereotactic radiosurgery End-To-End Phantom.
Conclusions:
We show that the novel MCCAO method generates VMAT plans that meet clinical objectives for both conventional and MR-linacs.
More Related Videos
10:23Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Mechanistic Models: Compartment Models in Algorithms for Numerical Problem Solving
In individual population analyses, different algorithms are employed, such as Cauchy's method, which uses a...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiation: Applications
The average...