Background:
For normothermic ex vivo heart perfusion (EVHP), a resting mode and working mode have been proposed. We newly developed a left ventricular assist device (LVAD) mode that supports heart contraction by co-pulse synchronized LVAD.
Methods:
Following resting mode during time 0 to 1 hour, pig hearts (n = 18) were perfused in either resting, working, or LVAD mode during time 1 to 5 hour, and then myocardial function was evaluated in working mode at 6 hour. The preservation ratio was defined as the myocardial mechanical function at 330 minute divided by the function at 75 minute. In LVAD mode, LVAD unloaded the pressure and the volume in the left ventricle in the systolic phase.
Results:
The LVAD group was significantly associated with higher preservation ratios in cardiac output (resting, 33 ± 3; working, 35 ± 5; LVAD, 76% ± 5%; p < 0.001), stroke work, dP/dt maximum, and dP/dt minimum compared with the other groups. Glucose consumption was significantly reduced in the resting group. The LVAD group was significantly associated with higher myocardial oxygen consumption (resting, 2.2 ± 0.3; working; 4.6 ± 0.5; LVAD, 6.1 ± 0.5 mL O2/min/100 g, p < 0.001) and higher adenosine triphosphate (ATP) levels (resting, 1.1 ± 0.1; working, 0.7 ± 0.1; LVAD, 1.6 ± 0.2 μmol/g, p = 0.001) compared with the others.
Conclusion:
These data suggest that myocardial mechanical function was better preserved in LVAD mode than in resting and working modes. Although our data suggested similar glycolysis activity in the LVAD and working groups, the higher final ATP in the LVAD group might be explained by reduced external work in LVAD.