Enhanced Depth of Focus Intraocular Lenses: Through Focus Evaluation of Wavefront-Shaping versus Diffractive Optics
Ruediger Schmid1, Andreas F Borkenstein2
1Accuratis, Practice for Refractive Eye Surgery, Ulm, Germany.
Introduction:
A new class of nondiffractive, wavefront-shaping Enhanced-Depth-of-Focus (EDoF) IOLs has been introduced very recently to cope with photic phenomena known from diffractive EDoF IOLs. We investigated the through focus modulation transfer function (MTF) of two wavefront-shaping EDoF IOLs compared to an established diffractive EDoF IOL on the optical bench. Such comparison on the optical bench had not been performed before and is of high clinical importance for the cataract surgeon.
Material And Methods:
Tecnis Symfony (diffractive) and the wavefront-shaping Acrysof IQ Vivity and LuxSmart Crystal IOLs (22 D each) were assessed by the OptiSpheric IOL PRO 2 imaging test bench with an ISO-2 cornea and a wavelength of 546 nm. Apertures of 3 mm and 4.5 mm were applied.
Results:
For all three IOLs, two peaks showed up in the through focus MTF curves representing the primary and secondary focus. For Symfony, these peaks were most distinct. Power difference between far and intermediate focus was 1.25 D (Symfony), 1.75 D (Vivity), and 1.5 D (LuxSmart) with an aperture of 3 mm. With an aperture of 4.5 mm, only for LuxSmart, power difference diminished slightly to about 1 D, and only the MTF in the intermediate focus decreased for all lenses.
Conclusion:
For all three IOLs, we could confirm a considerable depth of focus that was most extended for Vivity. Both new wavefront-shaping IOLs had lower values of peak MTF but a markedly more continuous through focus behavior compared to the diffractive EDoF IOL.
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
07:14Microfabrication of Implantable Optics Integrated in a Microstructured Imaging Window for Advanced In Vivo Imaging
Published on: April 11, 2025
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Confocal Fluorescence Microscopy
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...