DOSIMETRIC VERIFICATION OF INTENSITY MODULATED RADIOTHERAPY (IMRT) TREATMENT PLANS FOR PROSTATE CANCER PATIENTS
Iva Mrčela1,2, Marin Gregov1,2, Ante Matanić1,2
1Department of Medical Physics.
Acta Clinica Croatica
|March 20, 2023
Summary
Intensity modulated radiotherapy (IMRT) ensures accurate prostate cancer treatment dose delivery. Patient-specific quality assurance (PSQA) using advanced phantoms confirms treatment safety and accuracy.
Area of Science:
- Radiation Oncology
- Medical Physics
Background:
- Intensity modulated radiotherapy (IMRT) is a standard for localized prostate cancer.
- Transitioning to IMRT necessitates rigorous dose delivery verification.
- Conformal radiotherapy (3D CRT) preceded the adoption of IMRT.
Purpose of the Study:
- To present the clinical workflow for patient-specific quality assurance (PSQA) in IMRT for prostate cancer.
- To evaluate the accuracy of dose delivery in IMRT for prostate cancer patients.
- To assess the safety and compliance of IMRT treatments with established tolerance limits.
Main Methods:
- Utilized step-and-shoot IMRT for 40 prostate cancer patients.
- Performed dosimetric verification using a rotational phantom and Octavius 4D detector system.
- Conducted a three-dimensional comparison of calculated and delivered radiation dose distributions.
Main Results:
- All patient-specific quality assurance (PSQA) procedures demonstrated compliance with universal tolerance limits.
- The implemented IMRT workflow ensures safe and accurate radiation dose delivery.
- Results support the potential for implementing stricter quality assurance constraints in the future.
Conclusions:
- Patient-specific quality assurance (PSQA) is crucial for verifying IMRT accuracy in prostate cancer treatment.
- The study confirms the safety and reliability of the IMRT technique for prostate cancer.
- Findings pave the way for refining quality assurance protocols and potentially adopting more stringent dose delivery criteria.