Failure of a Highly Cross-Linked Polyethylene Liner After Spine Fusion
Kimberly Roesler1, Kevin L Garvin
1From the Creighton University School of Medicine, Omaha, NE (Dr. Roesler), and the Department of Orthopaedic Surgery, University of Nebraska Medical Center, Omaha, NE (Dr. Garvin).
Abstract:
A 73-year-old woman, 11 years after total hip arthroplasty and 2 years after a multilevel lumbar spine fusion, presented with a 2-month history of anterior hip and gluteal pain. She was diagnosed with an acetabular liner fracture that occurred through the high wall, likely related to repetitive impingement on the neck of the femoral implant, as notable burnishing was observed on the explanted femoral head. The acetabulum was successfully revised to a dual-mobility articulation. Spinal fusion after total hip arthroplasty can alter the acetabular implant position and was seen in our patient whose previously functional high-walled liner failed. Surgeons may consider alternative techniques, including a change in acetabular implant's anteversion to mitigate the need for a high-walled liner or the use of a dual-mobility bearing.