Related Experiment Videos
Determination of thiocyanate in plasma and saliva without deproteinisation and its validation as a smoking parameter
P Degiampietro1, E Peheim, D Drew
1Chemisches Zentrallabor der Universitätskliniken, Inselspital Bern.
Abstract:
A simple photometric procedure was developed for the determination of thiocyanate (SCN-) in plasma and saliva without deproteinisation or dialysis. Fe(III) ions form a red coloured complex with SCN- with a maximum absorbance at 460 nm. Mercury(II) nitrate is used to run a sample blank. A manual and an automated version (COBAS BIO) of the method is described. The method is linear up to 5000 mumol/l SCN-. The CV of the between-run precision is 2.8-8% for the manual and 2.6-6.6% for the automated method. The SCN- plasma concentration was 21-134 mumol/l in nonsmokers and 44-260 mumol/l in smokers. In mixed saliva, the concentration is much higher than in plasma: 1.57-5.5 mumol/l in smokers and 0.79-3.9 mumol/l in nonsmokers. Plasma, but not oral fluid SCN-, is a valuable parameter for studying smoking habits in population surveys. However, its use is limited. In our experience only heavy smokers can be distinguished from nonsmokers.