Validation of Monte Carlo simulated absorbed-dose-to-water inside a custom SPECT/CT phantom using active and passive
Andrew Bertinetti1, Miesher Rodrigues2, Benjamin Palmer1
1University of Wisconsin-Madison, United States of America.
Abstract:
Objective.This project aims to provide a novel method for performing dosimetry measurements on TRT radionuclides using a custom-made SPECT/CT compatible phantom, common active and passive detectors, and Monte Carlo simulations. In this work we present a feasibility study using99mTc for a novel approach to obtaining reproducible measurements of absorbed-dose-to-water from radionuclide solutions using active and passive detectors in a custom phantom for the purpose of benchmarking Monte Carlo-based absorbed-dose-to-water estimates.Approach. A cylindrical, acrylic SPECT/CT compatible phantom capable of housing an IBA EFD diode, SNC600c Farmer type ion chamber, and TLD-100 microcubes was designed and built for the purpose of assessing internal absorbed-dose-to-water at various points within a solution of99mTc. The phantom is equipped with removable inserts that allow for numerous detector configurations and is designed to be used for verification of SPECT/CT-based absorbed-dose estimates with traceable detector measurements at multiple locations. Three experiments were conducted with exposure times ranging from 11 to 21 h with starting activities of approximately 10-16 GBq. Measurement data was compared to Monte Carlo simulations using the egs_chamber user code in EGSnrc 2019.Main results. In general, the ionization chamber measurements agreed with the Monte Carlo simulations withink= 1 uncertainty values (±4% and ±7%, respectively). Measurements from the TLDs yielded results withink= 1 agreement of the MC prediction (±6% and ±5%, respectively). Agreement withink= 1 uncertainty (±6% and ±7%, respectively) was obtained for the diode for one of three conducted experiments.Significance. While relatively large uncertainties remain, the agreement between measured and simulated absorbed-doses provides proof of principal that dosimetry of radionuclide solutions with active detectors may be performed using this type of phantom with potential modifications for beta-emitting radionuclides to be introduced in future work.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021