Comparison of Efficiency and Effectiveness Between a Slap Hammer and Rigid C-Frame Device for Extraction of an
Adam Gosik-Wolfe1, Suvleen K Singh2, Megan E Welsh2
1Shukla Medical, St. Petersburg, FL; and.
Background:
Removal of orthopaedic intramedullary implants can be difficult and time-consuming. Instrumentation for implant removal is frequently deficient for effective removal. The purpose of this study was to compare the efficiency of a C-type jig with a standard slap hammer attachment. We hypothesize that a C-type jig will be a more energy-efficient method for implant removal.
Methods:
An intramedullary (IM) nail removal was simulated in a series of 10 tests using 40 PCF Sawbones bone blocks with drilled holes and custom-made IM nails. Each attachment was secured to a Shukla Medical threaded connector from their IM nail revision product. A camera recorded each hammer swing, and a caliper recorded the distance the nail traveled out of the bone block. The data were then analyzed to determine extraction rate and efficiency.
Results:
The c-frame hammer exerted a greater force, had a greater extraction efficiency, and required 37.4% less energy expenditure than the slap hammer to extract the nail the same distance. The c-frame hammer also removed the nail 38.1% faster with the same energy expenditure and possessed greater usable kinetic energy, whereas the slap hammer had more "lost" energy.
Conclusions:
The c-frame hammer attachment was found to have a considerably higher extraction rate and efficiency than the slap hammer. It will be a more useful method of implant extraction, especially for cases involving larger bones or larger implants. However, the slap hammer may be more suitable for smaller tools or bones for which larger impact loading would be detrimental.
More Related Videos
04:42Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
06:59Fracture Apparatus Design and Protocol Optimization for Closed-stabilized Fractures in Rodents
Published on: August 14, 2018
Related Concept Videos
Moment of a Force: Problem Solving
Mechanical Efficiency of Real Machines
However, in reality, no machine can be truly ideal, and all of them experience some...
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
Non-destructive Tests for Concrete Strength