The first mobile photon-counting detector CT: the human images and technical performance study
Su-Jin Park1, Junyoung Park1, Doil Kim1
1Health & Medical Equipment Business, Samsung Electronics Co., Ltd, 129, SamSung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16677, Republic of Korea.
Physics in Medicine and Biology
|March 29, 2023
Summary
Photon-counting-detector CT (PCD-CT) offers higher spatial resolution and multi-spectral imaging capabilities. This new technology demonstrates diagnostic equivalence to conventional scanners and enables advanced material decomposition for enhanced medical imaging.
Area of Science:
- Medical Imaging
- Radiology
- Biomedical Engineering
Background:
- Conventional energy-integrating detector CT (EID-CT) has limitations in spatial resolution and spectral information.
- Photon-counting-detector CT (PCD-CT) technology offers potential advancements in image quality and quantitative analysis.
- The mobile PCD-CT system (OmniTom Elite) has received FDA 510(k) clearance for clinical use.
Purpose of the Study:
- To evaluate the human imaging capabilities of a mobile PCD-CT system.
- To assess the 'on demand' higher spatial resolution and multi-spectral imaging features of PCD-CT.
- To demonstrate the feasibility and performance of PCD-CT in clinical settings.
Main Methods:
- Utilized an FDA-cleared mobile PCD-CT scanner (OmniTom Elite).
- Imaged certified CT phantoms and a human cadaver head to assess high-resolution (HR) and multi-energy imaging.
- Conducted first-in-human imaging on three volunteers to demonstrate clinical performance.
Main Results:
- First human PCD-CT images at 5 mm slice thickness were diagnostically equivalent to conventional EID-CT.
- PCD-CT HR mode achieved 11 lp/cm resolution, significantly higher than EID-CT's 7 lp/cm (standard mode).
- Quantitative multi-energy CT performance showed accurate material decomposition (iodine, calcium, water) with low error (3.25% in VMI).
Conclusions:
- PCD-CT enables multi-resolution acquisition without hardware changes, offering superior spatial resolution over conventional mobile EID-CT.
- The quantitative spectral capability of PCD-CT allows for accurate, simultaneous multi-energy imaging for material decomposition and VMI generation.
- PCD-CT represents a significant advancement in mobile CT technology, enhancing diagnostic capabilities and quantitative analysis.