Related Experiment Video
Updated: Aug 5, 2025

Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Commissural Alignment With ACURATE neo2 Valve in an Unselected Population
Christopher U Meduri1, Andreas Rück2, Rickard Linder1
1Karolinska University Hospital, Department of Cardiology, Stockholm, Sweden.
Commissural alignment during transcatheter aortic valve replacement (TAVR) with the ACURATE neo2 valve is feasible and safe. This technique achieved successful alignment in nearly all patients, improving coronary access and potentially valve durability.
Area of Science:
- Cardiology
- Interventional Cardiology
- Biomedical Engineering
Background:
- Commissural alignment in transcatheter aortic valve replacement (TAVR) is crucial for enhancing coronary access, future valve interventions, and valve longevity.
- The effectiveness of commissural alignment using the ACURATE neo2 prosthetic heart valve in a large patient cohort remains under investigation.
Purpose of the Study:
- To evaluate the feasibility and success rate of achieving commissural alignment in a broad TAVR patient population receiving the ACURATE neo2 valve.
- To assess the safety and procedural outcomes associated with this dedicated alignment technique.
Main Methods:
- A cohort of 170 consecutive patients undergoing TAVR with the ACURATE neo2 valve were implanted using a specific technique for commissural alignment.
- Valve orientation was adjusted by rotating the unexpanded valve within the aortic root, guided by right-left overlap and three-cusp fluoroscopic views.
- Effectiveness was quantified by comparing post-procedure fluoroscopic valve orientation with pre-procedural computed tomography cusp orientation, while safety was monitored for mortality, stroke, and other complications within 30 days.
Main Results:
- Out of 170 patients, 167 were analyzed for alignment, with 97% achieving successful alignment (mild or less misalignment).
- Specifically, 80% of patients demonstrated optimal commissural alignment.
- Minor degrees of misalignment were observed in 17% (mild), 1.2% (moderate), and 1.8% (severe) of cases, with no significant safety concerns reported.
Conclusions:
- This extensive study demonstrates that commissural alignment with the ACURATE neo2 valve is achievable in the vast majority of TAVR patients.
- The technique proved safe, with no adverse impact on procedure duration or increased safety risks.
- Commissural alignment using this novel method is both effective and safe for TAVR procedures.