Head-Mounted Display for Clinical Evaluation of Neck Movement Validation with Meta Quest 2
Manuel Trinidad-Fernández1,2, Benoît Bossavit3, Javier Salgado-Fernández4
1Grupo de Investigación Clinimetría, Departamento de Fisioterapia, Universidad de Málaga, 29071 Málaga, Spain.
Sensors (Basel, Switzerland)
|March 30, 2023
Summary
The Meta Quest 2 head-mounted display (HMD) accurately measures neck rotation angles for screening purposes. This virtual reality system offers a valid, low-error alternative for assessing neck mobility in healthy individuals.
Area of Science:
- Biomedical Engineering
- Rehabilitation Technology
- Virtual Reality Applications
Background:
- Neck disorders are prevalent and impactful.
- Head-mounted displays (HMDs) offer immersive virtual reality (iVR) experiences.
- Accurate screening tools for neck mobility are needed.
Purpose of the Study:
- To validate the Meta Quest 2 HMD as a tool for screening neck movement in healthy individuals.
- To assess the accuracy of neck angle measurements using HMD technology.
- To compare HMD-derived neck mobility data against a standard measurement system.
Main Methods:
- Developed a virtual reality application for participants to perform six neck movements.
- Collected head position and orientation data using the Meta Quest 2 HMD.
- Utilized an InertiaCube3 inertial measurement unit (IMU) as a criterion standard for comparison.
- Calculated mean absolute error (MAE), percentage error (%MAE), and performed Bland-Altman analysis.
Main Results:
- Average absolute errors for neck angles were below 1° (0.48 ± 0.09°).
- Rotational movement showed an average %MAE of 1.61 ± 0.82%.
- Head orientation correlations ranged from 0.70 to 0.96, with good agreement between HMD and IMU.
Conclusions:
- The Meta Quest 2 HMD system provides valid measurements of neck rotational angles across three anatomical axes.
- The system demonstrates acceptable error percentages and minimal absolute error for neck rotation.
- The Meta Quest 2 HMD is a viable sensor for screening neck disorders in healthy populations.