Related Experiment Video
Updated: Aug 5, 2025

Evaluation of Fluid Overload by Bioelectrical Impedance Vectorial Analysis
Published on: August 17, 2022
Elastic Textile Wristband for Bioimpedance Measurements
Giuseppina Monti1,2, Federica Raheli3, Andrea Recupero1
1Department of Engineering for Innovation, University of Salento, 73100 Lecce, Italy.
Abstract:
In this paper, wristband electrodes for hand-to-hand bioimpedance measurements are investigated. The proposed electrodes consist of a stretchable conductive knitted fabric. Different implementations have been developed and compared with Ag/AgCl commercial electrodes. Hand-to-hand measurements at 50 kHz on forty healthy subjects have been carried out and the Passing-Bablok regression method has been exploited to compare the proposed textile electrodes with commercial ones. It is demonstrated that the proposed designs guarantee reliable measurements and easy and comfortable use, thus representing an excellent solution for the development of a wearable bioimpedance measurement system.