Related Experiment Video
Updated: Aug 14, 2026

Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Deformable Sensors for Pressure and Position Assessment Using Time-Domain Reflectometry in Motor Rehabilitation
Andrea Cataldo1, Antonio Masciullo1, Giuseppina Monti1
1Department of Engineering for Innovation, University of Salento, 73100 Lecce, Italy.
Abstract:
This work presents the design and preliminary experimental validation of deformable sensors based on time-domain reflectometry (TDR) for rehabilitation-oriented interaction monitoring. Three architectures were investigated: a planar multilayer sensor and two coaxial configurations based on foam and engineered TPU-Hilbert structures. Controlled indentation tests were performed at different positions and deformation levels, extracting two TDR-derived features: the minimum reflection coefficient ρmin, related to deformation intensity, and the perturbation time tpert, related to contact localization. Preliminary calibration curves and two-dimensional maps were used to analyze the coupled dependence of the response on position and indentation depth. Application-oriented manual tests confirmed the different suitability of the three geometries for localized finger pressing, distributed two-hand grasping, and controlled single-hand squeezing. Overall, the results support TDR-based deformable sensors as low-complexity and geometry-adaptable tools for spatially resolved monitoring of motor rehabilitation interactions.