Related Experiment Video
Updated: Aug 4, 2025

Biochemical Measurement of Neonatal Hypoxia
Published on: August 24, 2011
Heart Disease Screening and False Hypoxemia in the Neonate
David Rabiço-Costa1, Laura Leite-Almeida1, Sara Geraldes Paulino1
1Pediatric Department, São João University Hospital Center, Porto, Portugal.
Background:
Screening for congenital heart diseases by pulse oximetry is used for the initial assessment of the neonate. Variants of hemoglobin F can compromise light absorbance, inducing erroneous results.
Case Report:
Two infants screened for congenital heart disease showed an asymptomatic low peripheral oxygen saturation. Arterial blood gases analysis revealed a normal arterial pressure of oxygen and oxygen saturation. More likely and/or severe causes of hypoxemia were ruled out. This "artifact" with SpO2-SaO2 dissociation, and after exclusion of other common etiologies of hypoxemia, raised the clinical suspicion of hemoglobinopathy. Hemoglobin molecular and genetic studies identified specific mutations in gamma chains from hemoglobin F, named hemoglobin F Sardinia.
Conclusion:
Hemoglobin F variants may result in low peripheral oxygen saturation readings by pulse oximetry, explaining the discordance in the clinical appearance and low peripheral oxygen saturation readings.
Related Concept Videos
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...