Related Experiment Video
Updated: Aug 4, 2025

13:35
Transcranial Direct Current Stimulation and Simultaneous Functional Magnetic Resonance Imaging
Published on: April 27, 2014
21.9K
Transcranial direct current stimulation with functional magnetic resonance imaging: a detailed validation and
Davide Nardo1,2, Megan Creasey3, Clive Negus3
1MRC Cognition and Brain Sciences Unit, University of Cambridge, Cambridge, UK.
Wellcome Open Research
|April 4, 2023
Summary
This study presents a safe and reliable method for combining transcranial direct current stimulation (tDCS) with functional magnetic resonance imaging (fMRI). The developed procedure ensures high-quality data for investigating brain function and recovery, such as after stroke.
Area of Science:
- Neuroscience
- Medical Imaging
- Brain Stimulation
Background:
- Transcranial direct current stimulation (tDCS) modulates brain activity for research and clinical use.
- Combining tDCS with functional magnetic resonance imaging (fMRI) allows direct investigation of brain region causality.
- Technical challenges and variability hinder the comparability and replicability of tDCS-fMRI studies.
Purpose of the Study:
- To develop and validate a reliable tDCS-fMRI method at 3 Tesla for studying language recovery post-stroke.
- To address technical challenges and methodological variability in tDCS-fMRI applications.
- To provide a Standard Operating Procedure (SOP) to promote replication and enhance data quality.
Main Methods:
- Developed a tDCS-fMRI protocol at 3T.
- Validated the method using a healthy volunteer in three conditions: No-tDCS, Sham-tDCS, and Anodal-tDCS.
- Analyzed MR data using region-of-interest (ROI) analyses, modeling scanner drift, motion, and temporal autocorrelation.
Main Results:
- No visible signal dropouts or distortions were observed due to tDCS equipment.
- Functional MR sensitivity remained unaffected across conditions in grey matter and ROIs.
- The tDCS set-up and stimulation protocol did not degrade MR data quality.
Conclusions:
- The developed tDCS-fMRI procedure is technically safe and reliable for acquiring high-quality MR data.
- The detailed SOP facilitates adoption and replication in future studies.
- This approach enhances the quality and reliability of tDCS-fMRI research.
Keywords:
fMRIfunctional magnetic resonance imagingimplementation guidesafety factorsstandard operating proceduretechnical challengestranscranial direct current stimulationtranscranial electrical brain stimulation