A 220 GHz-1.1 THz continuous frequency and polarization tunable quasi-optical electron paramagnetic resonance
Gyan Khatri1, Gregory Fritjofson1, Jacob Hanson-Flores1
1Department of Physics, University of Central Florida, Orlando, Florida 32816, USA.
Abstract:
Here, we describe a custom-designed quasi-optical system continuously operating in the frequency range 220 GHz to 1.1 THz with a temperature range of 5-300 K and magnetic fields up to 9 T capable of polarization rotation in both transmitter and receiver arms at any given frequency within the range through a unique double Martin-Puplett interferometry approach. The system employs focusing lenses to amplify the microwave power at the sample position and recollimate the beam to the transmission branch. The cryostat and split coil magnets are furnished with five optical access ports from all three major directions to the sample sitting on a two-axes rotatable sample holder capable of performing arbitrary rotations with respect to the field direction, enabling broad accessibility to experimental geometries. Initial results from test measurements on antiferromagnetic MnF2 single crystals are included to verify the operation of the system.
Related Concept Videos
UV–Vis Spectroscopy: Molecular Electronic Transitions
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....