Evaluation of the ability of a commercially available cuffless wearable device to track blood pressure changes
Isabella Tan1,2, Sonali R Gnanenthiran1,3, Justine Chan4
1The George Institute for Global Health.
Objectives:
Cuffless wearable blood pressure (BP) devices may allow detailed evaluation of BP for prolonged periods, but their ability to accurately track BP changes is uncertain. We investigated whether a commercially available cuffless wearable device tracks: 24-h systolic (SBP) and diastolic BP (DBP) compared to conventional ambulatory monitoring (ABPM); and antihypertensive medication-induced BP changes compared to cuff-based home BP monitoring (HBPM).
Methods:
We fitted 41 participants (32% females, 58 ± 14 years, 80% hypertensive) with a wrist-wearable cuffless BP device (Aktiia) continuously for 6-12 days. At the beginning and the end of this period, 24-h ABPM was performed. Three participants with hypertension (one female; 60 ± 8 years) wore the Aktiia device and performed HBPM continuously one week before and 2 weeks after antihypertensive medication uptitration.
Results:
Compared to ABPM, Aktiia reported higher average SBP for 24-h (difference 4.9 mmHg, 95% CI [1.9, 7.9]) and night-time (15.5[11.8, 19.1] mmHg; all P ≤ 0.01), but similar daytime (1.0 [-1.8, 3.8] mmHg; P = 0.48). Similarly, average cuffless DBP was higher for 24-h (4.2 [2.3, 6.0] mmHg) and night-time (11.8 [9.5, 14.1] mmHg; both P < 0.001), but similar during daytime (1.4 [-0.4, 3.23] mmHg; P = 0.13). Aktiia also reported reduced night-time dip for SBP (difference 14.2 [12.1, 16.3] mmHg) and DBP (10.2 [8.5, 11.9] mmHg; both P < 0.001). The average medication-induced SBP/DBP decline after 2 weeks of uptitration was -1.0/-0.8 mmHg with Aktiia vs. -19.7/-11.5 mmHg with HBPM ( P = 0.03 for difference).
Conclusion:
This cuffless wearable device did not accurately track night-time BP decline and results suggested it was unable to track medication-induced BP changes.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure