Author Spotlight: Establishing a Practical and Cost-Effective Protocol for Corneal Sensitivity Testing in Clinical Settings
Published on: August 2, 2024
Clinical application of the Swiss Liquid Jet Aesthesiometer for corneal sensitivity measurement
Daniela S Nosch1, Emanuele Käser, Tobias Bracher1
1Institute of Optometry, University of Applied Sciences and Arts Northwestern Switzerland (FHNW), Olten, Switzerland.
Clinical Relevance:
Corneal sensitivity represents an important indicator for corneal health, its innervation and hence also for ocular disease. It is therefore of great interest from a clinical and research perspective to quantify ocular surface sensation.
Background:
The aim of this prospective cross-sectional cohort study was to clinically test the within-day and day-to-day repeatability of the new Swiss Liquid Jet Aesthesiometer, employing small droplets of isotonic saline solution for repeatability, and correlate with the Cochet-Bonnet aesthesiometer in a cohort of participants of two different age groups, based on participant feedback (psychophysical method).
Methods:
Participants were recruited from two equally, large age groups: group A (18-30 years) and group B (50-70 years). The inclusion criteria were healthy eyes, Ocular Surface Disease Index (OSDI) ≤ 13, and no contact lens wear. Mechanical corneal sensitivity threshold measurements with means of liquid jet and Cochet-Bonnet methods were carried out twice during two visits (a total of four measurements), with a stimulus temperature equal to or slightly above the ocular surface temperature.
Results:
Ninety participants completed the study (n = 45 per age group, average age in group A: 24.2 ± 2.94 years, group B: 58.5 ± 5.71 years). The coefficient of repeatability for the liquid jet method was 2.56 dB within visits and 3.61 between visits. For the Cochet-Bonnet method, it was 2.27 dB within visits and 4.42 dB between visits (Bland Altman with bootstrap analysis). Moderate correlation was observed between the liquid jet and the Cochet-Bonnet method (r = 0.540, p < .001, robust linear regression).
Conclusions:
Swiss liquid jet aesthesiometry offers a new examiner independent method for corneal sensitivity measurement with acceptable repeatability and moderate correlation with the Cochet-Bonnet aesthesiometer. It offers a large stimulus pressure range of 100-1500 mbar and a precision of 1 mbar. Stimulus intensity can be tuned more precisely and much smaller sensitivity fluctuations may be potentially detected.