Related Experiment Video
Updated: Aug 3, 2025

A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
Factors affecting the evaluation and use of a hemoglobin A2 method - Lot-to-lot variation, long-term deviation and
Britta Landin1, Christin Sisowath2, Annelie Strålfors2
1Dept. Clinical Chemistry, Karolinska University Hospital, SE-171 76 Stockholm, Sweden; Dept. Laboratory Medicine, Karolinska Institutet, SE-171 76 Stockholm, Sweden.
Background:
We compared two Hb A2 methods, high pressure liquid chromatography and capillary zone electrophoresis, at two occasions. In 2014 the methods showed good agreement, while in 2020 HPLC results were clearly higher than CE results. This finding prompted us to investigate the external quality assessment (EQA) outcome and our total patient results obtained by high pressure liquid chromatography over several years.
Methods:
The methods compared were Bio-Rad Variant II Beta-Thal Short Program (HPLC) and Sebia Capilllarys Flex Piercing (CE).
Results:
Our annual patient results obtained by HPLC increased significantly from 2014 to 2020. A similar trend was also seen in our EQA results. When patient results were grouped according to different reagent lots it became evident that method comparisons might be severely affected by lot-to-lot variation. We also noticed that samples analyzed with the HPLC method following a sample containing a high proportion of Hb E where prone to give falsely increased Hb A2 results.
Conclusion:
Standardization of the measurement of Hb A2 is urgently needed. Furthermore, the lot-to-lot variation must be minimized. While waiting for these improvements each laboratory ought to repeatedly evaluate their distribution of patient Hb A2 results.
Related Concept Videos
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Oxygen Transport in the Blood
Data Validation
Key parameters for method validation include:
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Factors Affecting Erythropoiesis
Several factors influence the erythrocyte production rate, with tissue oxygen level being among the most critical. Intense exercise or high altitudes can cause tissue hypoxia, which triggers the kidneys to release more erythropoietin (EPO) into the bloodstream.
EPO then...
Blood Transfusion
Blood Transfusion Overview
A blood transfusion is a medical procedure used to replace blood lost due to injury, surgery, or to treat conditions such as anemia or cancer. During a transfusion, donor blood is...