Cochlear implantation programming characteristics and outcomes of cochlear nerve deficiency
Xingmei Wei1, Simeng Lu1, Biao Chen1
1Key Laboratory of Otolaryngology Head and Neck Surgery, Department of Otorhinolaryngology Head and Neck Surgery, Beijing Tongren Hospital, Capital Medical University, Ministry of Education, 1, Dongjiaominxiang, Dongcheng District, Beijing, 100730, People's Republic of China.
Purpose:
Due to the specificity of cochlear implantation (CI) programming parameters and outcomes in cochlear nerve deficiency (CND) patients, this study aimed to investigate the correlation between programming parameters and outcomes and further compare the difference between normal and CND groups.
Methods:
Ninety (95 ears) CND patients (normal cochlea, 39; malformed cochlea, 56) and seventy-nine (81 ears) normal cochlea patients who underwent CI surgery with either Med-El or Cochlear devices were included. The programming parameters and outcomes evaluated by the questionnaires were collected and compared among the normal CND, malformed CND, and normal groups in the two device groups, and their correlation was analyzed.
Results:
In the CND group, a reduced stimulation rate, higher pulse width, and triphasic pulse were needed in some cases. The stimulus levels of the CND group were significantly higher than that of the normal group (p < 0.05), but the outcomes of the CND group were significantly worse than that of the normal group (p < 0.05), and the stimulus level was significantly correlated with the outcomes (p < 0.05). However, there was no difference between normal and malformed CND groups. The non-auditory response was observed in the CND group, especially the ones with malformations.
Conclusion:
The CI programming parameters of some CND patients need to be adjusted, and a slower stimulation rate and higher pulse width are required sometimes. CND patients need a higher stimulus level than normal patients but their outcomes are poorer. Non-auditory response should be noticed in CND patients during programming.