Related Experiment Video
Updated: Aug 3, 2025

Novel Diagnostics in Revision Arthroplasty: Implant Sonication and Multiplex Polymerase Chain Reaction
Published on: December 3, 2017
Risk of Recall Associated With Modifications to High-risk Medical Devices Approved Through US Food and Drug
Jonathan R Dubin1,2,3, Jonathan R Enriquez1,4,5, An-Lin Cheng6,7
1Department of Orthopedics, University of Missouri-Kansas City.
Importance:
High-risk medical devices approved by the US Food and Drug Administration (FDA) can undergo modifications to their original premarket approval (PMA) via 1 of 5 types of supplements. Only panel track supplements (approximately 1%) require clinical data for approval. The association between device modifications and risk to patient safety has not previously been analyzed.
Objective:
To determine the association between PMA supplements and the risk of any device recall and high-risk (class 1) recall.
Design, Setting, And Participants:
In this cohort study, the FDA database was queried for original devices approved via PMA from January 1, 2008, through December 31, 2019. Supplement and recall data were obtained for these devices from January 1, 2008, through December 31, 2021, giving a minimum 2-year follow-up after initial approval. Data were analyzed from July 6 to August 6, 2022. Retrospective, time-to-event analysis investigated the association between the number and type of supplements and risk of recall.
Exposures:
Supplements submitted by manufacturers for FDA approval to modify devices.
Main Outcomes And Measures:
A mixed-effects Cox proportional hazards regression model with frailty terms was used, modeling device recall as an outcome variable during the observation period. A second model was performed for class 1 (high-risk) recall. Explanatory variables are the number of supplements, number of panel track supplements, and cardiovascular devices. Multivariable analysis was performed to identify independent risk factors for recall with hazard ratios (HRs) as the main end point.
Results:
A total of 373 original PMA devices with 10 776 associated supplements were included in the analysis. A median 2.5 (IQR, 1.2-5.0) supplements per device were approved annually. Cardiovascular devices contributed 138 supplements (37.0%), followed by microbiology with 45 (12.1%). No other specialty contributed more than 10%. Multivariable analysis demonstrated that each increase of 1 supplement per year was associated with increased risk of recall (HR, 1.28 [95% CI, 1.15-1.44]; P < .001). For class 1 recall, increased number of supplements (HR, 1.32 [95% CI, 1.06-1.64]; P = .01) and cardiovascular vsnoncardiovascular classification of devices (HR, 3.51 [95% CI, 1.15-10.72]; P = .03) were significantly associated with an increased risk of recall.
Conclusions And Relevance:
The findings of this cohort study suggest that PMA supplements are associated with an approximately 30% increased risk of any recall and class 1 recall. The FDA processes for approving modifications to high-risk medical devices should be reevaluated to optimize patient safety and public health.
Related Concept Videos
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...
Drug Regulation
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Guidelines and Strategies for Safe Computer Charting
Maintain Confidentiality and Security:
Clinical Trials: Overview
Desensitization and Tachyphylaxis