Poor comparability of plasma renin activity measurement in determining patient samples: the status quo and
Zhenni Liu1,2, Lizi Jin1,2, Jie Zeng1
1National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing, P.R. China.
Objectives:
This study aims to investigate and update the consistency and comparability of plasma renin activity (PRA) assays in measuring clinical samples. The contributions of recalibration, blank subtraction, and incubation strategies to interchangeability were also explored.
Methods:
Five different laboratories were evaluated using forty-six individual plasma samples, including four liquid chromatography-tandem mass spectrometry (LC‒MS/MS) assays and one chemiluminescence immunoassay (CLIA). Spearman correlation coefficient (R), Passing-Bablok regression, and Bland‒Altman plot analyses were used to evaluate the consistency among assays. Consistency before and after recalibration, blank subtraction, and incubation strategy unification was compared.
Results:
A good correlation was observed among all assays (R>0.93). None of the samples measured by all assays showed coefficient variation (CV) <10 %, and 37 % of samples showed overall CVs >20 %. The 95 % confidence intervals (CIs) for slopes did not contain 1 for most assay pairs. Large relative biases (-85.1-104.2 %) were found, and 76 % (52-93 %) of samples had unacceptable biases. Recalibration reduced the calibration bias. Ignoring blank subtraction improved the comparability across all assays while unifying incubation did not.
Conclusions:
The interchangeability of PRA measurement was unsatisfying. Harmonization on calibrator and ignoring blank were recommended. Unifying incubation strategy was unnecessary.
Related Concept Videos
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Renal Drug Clearance: Overview
Renal clearance can be calculated using different methods. One approach is to divide the urinary drug excretion rate by the plasma drug concentration. This method directly measures renal clearance, indicating the kidneys' efficiency in...
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Model-Independent Approaches for Pharmacokinetic Data: Noncompartmental Analysis
One important characteristic of noncompartmental analyses is that drug exposure increases proportionally with increasing doses. This...