Comparison of selected photon shield and organ-based tube current modulation for radiation dose reduction in head
Zilong Yuan1, Tao Liu1, Biao Zhang1
1Department of Radiology, Hubei Cancer Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
Objective:
The aim of this study is to investigate the radiation dose and image quality of head CT using SPS and OBTCM techniques.
Methods:
Three anthropomorphic head phantoms (1-yr-old, 5-yr-old, and adult) were used. Images were acquired using four modes (Default protocol, OBTCM, SPS, and SPS+OBTCM). Absorbed dose to the lens, anterior brain (brain_A), and posterior brain (brain_P) was measured and compared. Image noise and CNR were assessed in the selected regions of interest (ROIs).
Results:
Compared with that in the Default protocol, the absorbed dose to the lens reduced by up to 28.33%,71.38%, and 71.12% in OBTCM, SPS, and SPS+OBTCM, respectively. The noise level in OBTCM slightly (≤1.45HU) increased than that in Default protocol, and the SPS or SPS+OBTCM mode resulted in a quantitatively small increase (≤2.58HU) in three phantoms. There was no significant difference in CNR of different phantoms under varies scanning modes (p > 0.05).
Conclusions:
During head CT examinations, the SPS mode can reduce the radiation dose while maintaining image quality. SPS+OBTCM couldn't further effectively reduce the absorbed dose to the lens for 1-yr and 5-yr-old phantoms. Thus, SPS mode in pediatric and SPS+OBTCM mode in adult are better than other modes, and should be used in clinical practice.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Biological Effects of Radiation