The Preventiometer - reliability of a cardiovascular multi-device measurement platform and its measurement agreement
Martin Junge1,2, Markus Krüger3,4, Dietlind L Wahner-Roedler5
1Institute for Community Medicine, University of Greifswald, Greifswald, Germany.
Background:
Multimedia multi-device measurement platforms may make the assessment of prevention-related medical variables with a focus on cardiovascular outcomes more attractive and time-efficient. The aim of the studies was to evaluate the reliability (Study 1) and the measurement agreement with a cohort study (Study 2) of selected measures of such a device, the Preventiometer.
Methods:
In Study 1 (N = 75), we conducted repeated measurements in two Preventiometers for four examinations (blood pressure measurement, pulse oximetry, body fat measurement, and spirometry) to analyze their agreement and derive (retest-)reliability estimates. In Study 2 (N = 150), we compared somatometry, blood pressure, pulse oximetry, body fat, and spirometry measurements in the Preventiometer with corresponding measurements used in the population-based Study of Health in Pomerania (SHIP) to evaluate measurement agreement.
Results:
Intraclass correlations coefficients (ICCs) ranged from .84 to .99 for all examinations in Study 1. Whereas bias was not an issue for most examinations in Study 2, limits of agreement for most examinations were very large compared to results of similar method comparison studies.
Conclusion:
We observed a high retest-reliability of the assessed clinical examinations in the Preventiometer. Some disagreements between Preventiometer and SHIP examinations can be attributed to procedural differences in the examinations. Methodological and technical improvements are recommended before using the Preventiometer in population-based research.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure