Ultrasound Calculation of Fat Volume for Liposuction: A Clinical Software Validation
Background:
Fat manipulation procedures such as liposuction contain a degree of subjectivity primarily guided by the surgeon's visual or tactile perception of the underlying fat. Currently, there is no cost-effective, direct method to objectively measure fat depth and volume in real time.
Objectives:
Utilizing innovative ultrasound-based software, the authors aimed to validate fat tissue volume and distribution measurements in the preoperative setting.
Methods:
Eighteen participants were recruited to evaluate the accuracy of the new software. Recruited participants underwent ultrasound scans within the preoperative markings of the study area before surgery. Ultrasound-estimated fat profiles were generated with the in-house software and compared directly with the intraoperative aspirated fat recorded after gravity separation.
Results:
Participants' mean age and BMI were 47.6 (11.3) years and 25.6 (2.3) kg/m2, respectively. Evaluation of trial data showed promising results following the use of a Bland Altman agreement analysis. For the 18 patients and 44 volumes estimated, 43 of 44 measurements fell within a confidence interval of 95% when compared with the clinical lipoaspirate (dry) volumes collected postsurgery. The bias was estimated at 9.15 mL with a standard deviation of 17.08 mL and 95% confidence interval between -24.34 mL and 42.63 mL.
Conclusions:
Preoperative fat assessment measurements agreed significantly with intraoperative lipoaspirate volumes. The pilot study demonstrates, for the first time, a novel companion tool with the prospect of supporting surgeons in surgical planning, measuring, and executing the transfer of adipose tissues.