Sterile hypopyon secondary to ultrasonic cleaning solution
Journal of Cataract and Refractive Surgery
|May 1, 1986
Summary
Sterile hypopyon after cataract surgery was linked to endotoxin contamination on instruments cleaned in older ultrasonic solutions. This heat-stable endotoxin survived autoclaving, causing inflammation despite standard sterilization.
Area of Science:
- Ophthalmology
- Surgical Infections
- Sterilization Science
Background:
- A series of sterile hypopyon cases occurred post-cataract surgery.
- The cause was initially unknown despite thorough investigation of surgical protocols.
Purpose of the Study:
- To identify the source of sterile hypopyon following uncomplicated cataract surgery.
- To determine the link between ultrasonic cleaning solutions and postoperative inflammation.
Main Methods:
- Clinical observation of 21 sterile hypopyon cases post-cataract extraction.
- Investigation of surgical instruments and sterilization techniques.
- Laboratory analysis of instrument contamination using heat-stable endotoxin identification.
Main Results:
- A correlation was found between sterile hypopyon and the use of older ultrasonic cleaning solutions for surgical instruments.
- Instruments were contaminated with a heat-stable endotoxin that persisted through autoclaving.
- Contaminated instruments led to severe postoperative inflammatory reactions.
Conclusions:
- Older ultrasonic cleaning solutions can harbor heat-stable endotoxins.
- Endotoxin contamination on surgical instruments, even after autoclaving, can cause sterile hypopyon.
- Surgeons using ultrasonic cleaners must be aware of potential endotoxin contamination risks.