Validating scapular motion measurements using an optical motion analyzer and gravity magnetic resonance imaging
Yushin Mizuno1, Hiroichi Miaki2, Naoki Ohno2
1Section of Rehabilitation, Kanazawa University Hospital: 13-1 Takaramachi, Kanazawa, Ishikawa 920-8641, Japan.
Journal of Physical Therapy Science
|May 3, 2023
Summary
Scapular motion analysis using optical markers showed poor agreement with magnetic resonance imaging, questioning its validity. Further research is needed to confirm these findings on scapular kinematics.
Area of Science:
- Biomechanics
- Orthopedics
- Medical Imaging
Background:
- Accurate scapular motion measurement is crucial for diagnosing and treating shoulder pathologies.
- Optical motion analysis systems are commonly used to assess scapular kinematics.
- Multi-posture magnetic resonance imaging (MRI) offers a detailed view of anatomical structures and their movements.
Purpose of the Study:
- To validate scapular motion measurements obtained via optical motion analysis against those derived from multi-posture MRI.
- To assess the reliability of using pads with retroreflective markers for scapular motion tracking.
Main Methods:
- Twelve healthy males participated, with scapular angles measured at various shoulder flexion and abduction angles.
- Scapular angular changes (upward/downward, internal/external rotation) were calculated using an optical motion analyzer (VICON MX).
- Measurements were compared against scapular angles derived from multi-posture MRI data.
Main Results:
- A significant lack of agreement was observed between the optical marker and MRI-based scapular motion measurements in most tested conditions.
- No consistent bias was detected between the two measurement methods.
- The results indicate potential discrepancies in quantifying scapular kinematics using the evaluated optical system.
Conclusions:
- The validity of scapular motion analysis using pads with optical markers is questionable based on these findings.
- Limitations in the study environment may have influenced the results.
- Further validation studies are recommended to definitively assess the reliability of this optical measurement technique for scapular kinematics.