Polyethylene wear simulation models applied to a prosthetic hip joint based on unidirectional articulations
Donald Dalli1, Leonardo Fanton1, Bertram Mallia1
1Department of Metallurgy and Materials Engineering, Faculty of Engineering, University of Malta, Msida, MSD 2080, Malta.
Abstract:
Ultra-high molecular weight polyethylene (UHMWPE) is commonly used as soft-bearing material in total joint replacements. However, the release of polymeric wear debris is still related to complications leading to aseptic loosening. Recently, a novel hip prosthesis showing reduced wear was developed by the authors of this study, consisting of unidirectional cylindrical articulations instead of the conventional multidirectional ball-and-socket design. This study evaluates four different theoretical wear models applied to this new design. The calculated volumetric wear was compared to experimental results. Although all models provided a good indication of the wear rates for the ball-and-socket prosthesis, they exhibited high discrepancies when predicting the amount of wear of the new unidirectional design. It was observed that the closest agreement with experimental results was obtained by the models that consider the friction-induced molecular orientation phenomenon exhibited by UHMWPE.