Related Experiment Video
Updated: Jul 31, 2025

A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Two-screw osteosynthesis of the mandibular condylar head with different screw materials: a finite element analysis
Daphne Schönegg1, Andreas Koch2, Günter T Müller1
1Department of Oral and Maxillofacial Surgery, University Hospital Zurich, Zurich, Switzerland.
Abstract:
This study compared the biomechanical behavior of titanium, magnesium, and polylactic acid screws for two-screw osteosynthesis of mandibular condylar head fractures using finite element analysis. Von Mises stress distribution, fracture displacement, and fragment deformation were evaluated. Titanium screws performed the best in terms of carrying the highest load, resulting in the least fracture displacement and fragment deformation. Magnesium screws showed intermediate results, while PLA screws were found to be unsuitable with stress values exceeding their tensile strength. These findings suggest that magnesium alloys could be considered a suitable alternative to titanium screws in mandibular condylar head osteosynthesis.