Sodium concentration imaging in dermis layer by square-wave open electrical impedance tomography (SW-oEIT) with
Isnan Nur Rifai1,2, Marlin Ramadhan Baidillah1,3, Ridwan Wicaksono4
1Department of Mechanical Engineering, Graduate School of Science and Engineering, Chiba University, Chiba 263-8522, Japan.
This study introduces square wave open electrical impedance tomography (SW-oEIT) with spatial voltage thresholding (SVT) for imaging dermal sodium concentration. SW-oEIT with SVT provides a 15.32% higher correlation coefficient compared to conventional oEIT.
Area of Science:
- Biomedical Engineering
- Electrical Impedance Tomography
- Medical Imaging
Background:
- Accurate imaging of sodium concentration in the dermis is crucial for understanding physiological processes and diagnosing conditions.
- Conventional open electrical impedance tomography (oEIT) methods have limitations in resolving dermal sodium levels.
Purpose of the Study:
- To develop and validate a novel imaging technique, square wave open electrical impedance tomography with spatial voltage thresholding (SW-oEIT with SVT), for quantifying sodium concentration in the dermis.
- To compare the performance of SW-oEIT with SVT against conventional oEIT.
Main Methods:
- The SW-oEIT with SVT method involves three steps: voltage measurement using square wave current, spatial voltage thresholding to isolate the dermis, and sodium concentration imaging via Gauss-Newton reconstruction.
- The technique was applied to multi-layer skin simulations and ex-vivo experiments with varying sodium concentrations (5-50 mM).
- Image evaluation used the determination coefficient (R²) and normalized sensitivity (S) to assess the relationship between spatial mean conductivity and sodium concentration.
Main Results:
- SW-oEIT with SVT successfully imaged spatial mean conductivity distribution in the dermis, correlating with increasing sodium concentrations in both simulations and experiments.
- An optimized threshold distance (dΓ) of 2 mm yielded the highest evaluation values (R²=0.84, S=0.83).
- SW-oEIT with SVT demonstrated a 15.32% higher correlation coefficient (CC) compared to conventional sine wave oEIT.
Conclusions:
- SW-oEIT with SVT is an effective method for imaging dermal sodium concentration.
- The optimized SVT parameter (dΓ=2 mm) enhances imaging accuracy.
- This novel technique offers improved performance over conventional oEIT for sodium concentration imaging in biological tissues.